(51)
Int.Cl.6:
H
01 J 37/28
H
01 L 51/30
//
G01H 17/00
(19) Federal
Republic of Germany
[emblem]
German
Patent Office and Trademark Office
(12) Offenlegungsschrift
[=
published patent application]
(10) DE
198 22 634 A1
(21) Application
number: 198 22 634.9
(22) Filing
date: May 20,
1998
(43) Disclosure
date: November 25, 1999
|
(71) Applicant: Fuchs, Harald, Prof.Dr.,
48301 Nottuln, DE |
(72) Inventors: Fuchs, Harald, Prof.Dr.,
48301 Nottuln, DE; Simon, Ulrich, 45141 Essen, DE (56) Documents taken into consideration
for evaluating the patentability: WO 98 05 920 A1 HONGJIE DAI, et al.: Nanotubes
as nanoprobes in scanning probe microscopy. in: Nature, Vol. 384, 14 November
1996, pages 147-150. ARIE, Takayuki, et al.: Growth
of tungsten carbide nano-needle and its application as a scanning tunnelling
microscope tip. in: J. Phys. D: Appl. Phys. 31, 1998, pages L49 - L51; YOO, M.J., et al.: Scanning
Single Electron Transistor Microscopy: Imaging Individual Charges. in:
Science, Vol. 276, 25 April 1997, pp. 579-582. |
The following
information is derived from the documents submitted by the applicant.
(54) Nano-Atomic
Force Microscope with a Single Electron Transistor
DE 198 22 634 A1
Description
Prior
Art
The practical usability of atomic force microscopes
depends on two essential parameters:
1. The stiffness
(sensitivity) of the measurement probe (compliance) has to be chosen adequately
low (high) so that, when the measurement probe (tip) makes contact with the
surface to be examined, said surface is not destroyed. In the case of typical
bonding energies in solids of a few eV, this means an effective interatomic
spring constant ranging from 1 to 10 N/m (based on a simple spring model in two
dimensions and surface oscillation frequencies of 1012s-1).
Therefore, the stiffness of the measurement probe should be < 1 N/m.
2. The natural frequency of
the sensor (cantilever) should be chosen high in order to (a) guarantee fast
imaging of the surface and (b) achieve a large signal-to-noise ratio for
vibrations from the environment.
Owing to the non-linearly scaling ratio:
w = √ (E/z) (d/L2) (natural
frequency)
k =
(1/4) E (d4/L3) (spring constant,
square cross-section).
both conditions can be significantly improved by decreasing
the size of the structures (thickness d, length l). Modern sensors for atomic
force microscopy are made by means of micro-fabrication processes in order to
satisfy the two aforementioned requirements simultaneously. The fabrication
materials are usually Si, SiO2 and Si3N4.
Typical dimensions range from 100 to 200 micrometers (length), from 10 to 20
micrometers in width and 1 to 3 micrometers in thickness. The mechanical
natural frequencies of these micromechanical cantilevers ranges from 1 K Hz to
1 MHz.
A
significant increase in the natural frequencies must be expected if the linear
dimensions of the cantilevers are further reduced. A simple estimate shows that
cantilevers having a linear dimension in the nanometer range should reach
natural frequencies in the range of a few 100 HMz. The result was that
extremely fast and simultaneously soft cantilevers could be manufactured. Owing
to the high natural frequencies such cantilevers make it possible to measure
the surface faster and with higher sensitivity than is possible with the
currently available micromechanical cantilevers.
Approach
to a Solution
1.
Nano-Cantilevers
A
molecular dynamic simulation, designed by D. H. Robertson, Purdue University,
and T. White, NRL, showed that multi-wall carbon nano-tubes having a total
length of 300 mm and 12,000 atoms exhibited natural frequencies ranging from
1.1 to 1010 HZ (10 GHz). The outside diameter was 1.03 nm (citation:
see, for example, T.W.Ebbesen, 'Carbon Nanotubes', Physics Today, p. 26, June
1996). The use of such systems as AFM sensors (nano-AFM) would make it possible
to construct sensitive sensors for interatomic forces that are real
nano-sensors. The small tube diameter on the order of nanometers makes it
possible to use the tubes directly as the measurement tips. That is, there is
no need to modify the nano-cantilever with a measurement tip, as is required in
the case of micromechanical cantilevers. A slight tilt of the cantilever in
relation to the specimen surface suffices to bring the pointed end of the tip
in contact with the surface. The amplitudes of the cantilever are on the order
of a few nanometers. One end of the nano-cantilever can be mounted on a solid
substrate by spontaneous absorption (self assembly) or by targeted chemical
interaction of a suitably functionalized cantilever. This substrate can have
macroscopic dimensions (micrometers) and can be manufactured by conventional
methods, for example, from Si or similar materials.
2.
Nano-Cantilever Drive
Conventional
AFMs are either operated in static mode - i.e. in contact with the surface - or
in dynamic mode by means of an external drive, e.g. a piezo-actuator and/or are
excited in their natural resonance and/or higher harmonics by means of the
specimen. As Johansmann (A. Roters et al., J. Phys. Condens Matter, Vol. 8,
7561 - 7577, 1996) showed, instead of the external drive, thermal excitation of
soft cantilevers can also be used directly (fluctuations-dissipation-theorem).
Microscopy exploits the principle that the resonance conditions of the
cantilever change as it approaches the surface. Therefore, variations of the
amplitude, the phase and optionally the frequency are used. Whereas in
micromechanical cantilevers the thermally excited amplitudes are relatively
small (a feature that limits the application of this method), further
miniaturization of the sensors does result in relatively large amplitudes,
generated directly by thermal excitation. Therefore, if a nano-cantilever, like
the above described nano-tubes, is used, there is absolutely no need for an
external drive for the dynamic mode. The high mechanical natural frequency
(typically from 1 to 10 GHz), which would be generated by external means only
with difficulty, is produced directly as the molecular deflection oscillation.
It is a function of the size of the nano-tube and can be varied by chemical
modification, such as embedding metal or metal oxide nano-particles. The
results are totally new considerations for measuring interatomic and/or
intermolecular forces.
3.
Readout
In order
to read out the sensor signal, conventional atomic force microscopes with
micromechanical force sensors use optical, capacitive or tunnel transitions.
The most wide-spread are optical light spots. However, these methods cannot be
used owing to the small dimensions of the nano-tubes, since an adequate
signal-to-noise ratio cannot be achieved with these methods. One exception is
electron tunnelling. This approach shall be pursued in the method proposed
here.
a)
(Quasi) Static Operation
When
semiconducting and/or metallic cathode ray tubes are used, it is possible to
implement a quasi-static operation with the use of conductive specimens. When
the sensor approaches the conductive surface, the result is a tunnel contact
(< 0.4 nm: point contact) at intervals of < 1 nm. A measurable tunnel
current can flow over this tunnel contact. The intensity of this current can be
regarded (assuming a constant charge density of the surface) as the distance
gauge and, thus, as the topography signal.
b)
Dynamic Operation
The
nano-AFM is especially suitable for a dynamic operating mode, since it does not
require any external mechanical source of excitation for the dynamic operating
mode (direct excitation by means of thermal fluctuations). However, for the
so-called non-contact operating mode, the tunnel contact over a conductive
surface can no longer be used. One solution to this problem is to use fast
amplifiers, which detect directly the local deflection of the cantilever. The
conventional current amplifiers that are used for electron tunnelling
microscopy are ruled out, because high gains (1010) with FET /
bipolar technology are associated with large time constants (RC constant).
Suitable are amplification methods that allow a direct resolution and
measurement of single electrons when an oscillating measurement probe
approaches the readout sensor.
As a
solution to the problem, we propose a readout sensor on the basis of single
electron transistors (English: single electron transistor, SET). Owing to the small
characteristic intrinsic capacities (typically 1018 F), these
quantum components permit extremely fast switching actions and the direct
detection of single electrons. According to the invention, the steering element
shall be the nano-cantilever, which approaches (thermally excited) in an
oscillating manner the control gate of a suitable SET. In a suitable design the
upper or lower reversal point of the nano-cantilever, which is brought into
proximity with the sensor, shall be switched by one or more electrons in the
SET. The corresponding time constants are on an order of the oscillations of
the cantilever. Charged (polar) chemical groups that on approaching the gate of
the SET trigger the switching action and that are located on the end of the
nano-cantilever are suitable for enhancing the sensitivity of the sensor.
Suitable chemical groups can be attached on the end of the cantilever in the
aforementioned way by means of chemical functionalization of the cathode ray
nano-tubes. Another possibility for detection is to use fully conjugated linear
organic compounds having charged side groups, whose charge state and, thus, the
electric conductivity of the entire chain is affected by the presence of the
cantilever.
The
proposed method is presented as a schematic drawing in Figure 1. In this
drawing the reference numeral 1 denotes the nano-tube structure, one end of
which is clamped in a suitable suspension 7 and the left end of which can
oscillate freely in the direction of the arrow. Therefore, the bottom edge of
the nano-tube reaches the surface structure 4 that is to be measured and that
can be moved on a suitably scannable device, which can be scanned in the x and
y direction (see crossed arrows). The movement can be moved by means of a
nano-tube 8 in the nanometer range and below up to a few 100 micrometers. The
adjustment with the nano-tube 8 can be carried out in both the horizontal and
vertical direction. The sensor consists of the SET detector 6, which can be
built up with suitable semiconductor structures, e.g. nano-clusters. The SET
detector is coupled to the macroscopic environment by means of suitable
electrode structures 5. When the nano-tube moves oscillatingly, the single
electron detector in the vicinity of the upper reversal point of the tube is
steered by varying the potential, for example by means of a charged group 2.
Thus, an electron and/or a few electrons inside the electron structures 5
is/are moved each time, with the result that the oscillation is detected. The
detector (SET sensor) is brought into proximity with the oscillating nano-tube
by means of suitable measures, in particular piezo-ceramics.
Patent
Claims
1. Atomic force microscope comprising a cantilever and a
readout sensor, which can detect the deflections of the cantilever,
characterized by a single wall or multi-wall nano-tube as the cantilever.
2. Atomic force microscope, as claimed in claim 1,
characterized by a single electron transistor (SET) as the readout sensor.
3. Atomic force microscope, as claimed in claim 1 or 2,
characterized in that the free end area of the nano-tube on the side facing the
readout sensor exhibits charged (polar) chemical groups, whose approach is
detectable by the readout sensor, and that preferably when a single electron
transistor is used as the readout sensor, the approach of the chemical group
triggers a detectable switching action.
4. Nano-AFM, characterized in that, as claimed in claims
1 and 2, nano-tubes made of transition metal oxides, like VOx or BN,
are used.
5. SET, as claimed in claim 2, characterized by
components, comprising ligand-stabilized metal or semiconductor clusters, as
the SET element.
6. SET, as claimed in claim 2, characterized by SiO-SET
components as the SET element.
7. SET, as claimed in claim 2, characterized by
claso-carboranes or their derivatives as the SET element components.
8. SET, as claimed in claim 2, characterized in that
fullerenes are used.
------------------
1
page of drawings
------------------
-
Blank Sheet -